ACCU-CHEK INFORM II BLOOD GLUCOSE DETERMINATION
I. PURPOSE AND PRINCIPLE
The Accu-Chek Inform II meter is used for Quantitative determination of glucose in
whole blood specimens. The enzyme on the test strip, mutant variant of quinoprotein
glucose dehydrogenase converts the glucose in the blood sample to gluconolactone.
This reaction creates a harmless electrical current that the meter interprets as a glucose
concentration in mg/dL.
The system is calibrated with venous blood containing various glucose concentrations
and is calibrated to deliver plasma-like results. The reference values are obtained using
a validated test method. This test method is referenced to the hexokinase method and
is traceable to an NIST standard.
II. SCOPE
The Accu-Chek Inform II meter is used for monitoring blood glucose levels in hospitals,
ambulatory practices, and home care settings. It is used in both adult, pediatric, neonate
patient populations, however, it’s accuracy and precision in critically ill patients has not
been evaulated. The test is to be used for monitoring glucose levels and should not be
used for diagnosis.
III. PERSONNEL
Intended for use by clinical personnel who have received training and demonstrated
competency in this procedure. In the hospital setting, this includes Clinical Laboratory
Scientists, Registered Nurses, Nurse Practitioners, Physician Assistants, Physicians,
Respiratory Tech. In the ambulatory setting, this includes the aforementioned
personnel as well as Medical Assistants, Licensed Vocational Nurses and other
licensed Technologists.
IV. SPECIMEN REQUIREMENTS AND INTERFERENCES
A. Use capillary whole blood samples obtained from fingerstick. For children under
12 months of age, sample is obtained from heel stick. Fingerstick and heel stick
samples must be tested immediately. Do not use plasma, serum or Cord Blood
samples.
B. Wipe away the first drop of blood with gauze and use the second free-flowing
drop of blood from the fingerstick site for testing.
C. Venous, arterial or neonatal blood sample collected in lithium heparin are also
acceptable. Sample must be tested within 30 minutes of drawing to minimize the
effects of glycolysis.
Rev 1, Ver 2, 9/9/2013
Zobrazit stránku 0
1 2 3 4 5 6 ... 16 17

Shrnutí obsahu

Strany 1 - Testing

1 of 16Director Signature__________________________________ Date___________________Clinical LaboratoriesPoint of Care Testing ACCU-CHEK INFORM II

Strany 2

10 of 16D. Do not allow excess moisture to seep into the meter by squeezing off excess cleaning solution from the pre-moistened cleaning cloth before

Strany 3

11 of 16XVII. RECORDS MAINTENANCEA. Individual patient results are documented in the patient’s medical record, and also downloaded to the Electronic M

Strany 4 - Control is due. Required

12 of 16XX. TROUBLESHOOTINGA. On Screen MessagesB. Error CodesAll error messages displayed by the system have a letter identifying the message type, a

Strany 5 - “Control lot not found

13 of 16Display/Symptom/ERROR CODEPOSSIBLE SOLUTIONMeter displays “QC Due: Immediately”Run controls immediately.Patient testing cannot proceed until c

Strany 6

14 of 16C. Meter Reset Procedure1.Place the meter face down on a level surface.2.Press the reset button in the middle of the battery pack using a tool

Strany 7

15 of 16F. Troubleshooting Contact Information:1. For BROKEN METER, LOANERS, Base Unit, & Meter Battery: a. Contact CLINICAL ENGINEERING:1. Parna

Strany 8

16 of 16XXI. REFERENCESA. Internal PolicyB. Accu-Chek Inform II Blood Glucose Monitoring System Operator’s Manual 2013C. ACCU-CHEK Inform II Blood Glu

Strany 9

Signature ManifestDocument Number:SOP-0478Revision:1Title: Accu-Chek Inform II ProcedureAll dates and times are in Pacific Standard Time.Accu-Chek Inf

Strany 10 - Rev 1, Ver 2, 9/9/2013

2 of 16D. Alternate site capillary testing is unacceptable in a hospital setting. Due to the potential for less vascular circulation in alternate sit

Strany 11

3 of 163. Quality Control Solutionsa. Store the controls at room temperature at 15-30°C (59 - 86°F).b. Do not refrigerate or freeze. Unopened control

Strany 12

4 of 168. When you want to check the performance of the system.B. Results for Test Strip, Level 1 Low Control and Level 2 High Control Solutions shoul

Strany 13

5 of 16C. Press and release the “barcode” button. The button now appears with a black background during the scan.D. Hold the meter so that the window

Strany 14

6 of 16R. For custom comment, press icon to enter other or additional pertinent customized comment(s), not listed in the pre-programmed co

Strany 15

7 of 16Hold the meter so that the window of the barcode scanner is approximately 4 – 8 inches above the barcode on the patient’s ID band. The meter b

Strany 16 - REFERENCES

8 of 16c. Notified MD/NP d. Lab Drawe. For custom comment, press icon to enter other or additional pertinent customized comment(s), not li

Strany 17

9 of 16C. In case of a malfunction of the wireless transmission, there is a base unit wired to the network to download the meters on each nursing unit