Accu-Chek Go Servisní příručka Strana 172

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 208
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 171
172
Symbols, error messages and troubleshooting
Farbe: P 021 – Prüfmittelnummer: 201
This error message ... means:
O You bent or moved the test strip dur-
ing the test.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip.
O You inserted the test strip incorrectly
or did not press it right in.
Push the ejector down to eject the
test strip. Hold the test strip so the
arrows printed on it are uppermost
and pointing towards the test strip
guide. Without bending it, gently push
the test strip into the test strip guide
until you feel it click into place.
O The meter was exposed to a strong
electromagnetic field. Eject the test
strip or turn the meter off. Move
somewhere else or turn off the
source of the field.